Continuous analysis of CEMS emissions
Fields of application
Continuous monitoring of gas emissions from boilers, industrial furnaces, engines or waste incinerators.
Features
Fuji Electric France combines experience and expertise in the two related fields of robust, certified measurement of atmospheric emissions with Fuji CEMS, and standardized data management and reporting with Fuji DAHS. Fuji CEM Systemv7 is the latest evolution of the Fuji DAHS data management tool. It is a standardized acquisition and processing system for emission measurement data. It benefits from the latest technological advances in terms of digital, wireless, cloud-based and secure communications (for GSM), and the ability to send alerts by e-mail or SMS. It ensures strict compliance with current European regulations such as EN17255.
Measurable components | CO, CO₂, NOx, N₂O, SO₂, NH₃, HCl, CH₄, O₂, H₂O | |
Measurement type | Pressure, Temperature, Flow rate/Velocity, Dust content | |
Measurement principles | Non-dispersive infrared (NDIR) | CO, CO₂, NOx, N₂O, SO₂, CH₄ |
Zirconia, paramagnetic or electrochemical | O₂ | |
Differential zirconia measurement | H₂O | |
Laser | NH₃, HCl, H₂O | |
Optical LED diffusion (other methods optional) | Dust | |
Microventuri (other methods optional) | Flow | |
Capacitive | Pressure | |
PT100 temperature sensor | Temperature | |
Response time (t90) | < 200 s (NDIR) < 5 s (laser) | |
User interface | 10" touch screen with Fuji CEMS Managerv7 software module | |
Outputs / communication | Communication protocol | Modbus® TCP/IP as standard Options: 4-20 mA, Profibus, Modbus® RTU, Internet / GSM (FBOX) |
Reports | Automatic calibration in progress, maintenance, normal operation, controlled chimney | |
Alarms | General alarm, low flow, condensate, sampling probe, heated line, cooler, analysers | |
Contact outputs | Standard general alarm and optional specific alarms | |
Contact inputs | Boiler on (for multi-sampling mode and use of Fuji DAHS system) | |
Optional functions | Automatic calibration, automatic blowback Automatic QAL3 (combined with Fuji ACE DataQAL3 software module) | |
Multi-sampling | Up to 4 chimneys as standard. Up to 8 chimneys optional. | |
Sampling system | Sampling probe, heated line, filters, coolers and pumps depending on customer application: humidity, quantity and type of dust, pressure and flow rate, corrosion, stack diameter, etc. | |
Calibration gases (supplied locally) | Zero calibration | Nitrogen or clean, dry air |
Scale calibration | CO, CO₂, NO, SO₂, CH₄, N₂O, NH₃, HCl: 90% of each component measured in N₂ | |
Zirconium oxide | 21% O₂ in N₂ and 2% O₂ in N₂ | |
Power supply | 100, 110, 115, 200 or 230 Vac, 50/60 Hz | |
Dimensions (mm) | Interior installation cabinet for one chimney: 600 (W) x 800 (D) x 1800 (H) Other dimensions available according to application |
Type ofanalyser | ZRE: Single-beam NDIR | ZKJ: Dual-beam NDIR | ||
---|---|---|---|---|
NDIR Measurement scales (see available combinations on theanalyser data sheet) | NOx | 0 - 200 / 0 - 5000 ppm | NOx | 0 - 50 / 0 - 5000 ppm |
SO₂ | 0 - 200 / 0 - 5000 ppm | SO₂ | 0 - 50 / 0 - 5000 ppm | |
CO | 0 - 200 / 0 - 5000 ppm | CO | 0 - 50 / 0 - 5000 ppm | |
CH₄ | 0 - 500 / 0 - 5000 ppm | CH₄ | 0 - 200 / 0 - 5000 ppm | |
CO₂ | 0 - 10 ... 20 % | N₂O | 0 - 200 / 0 - 5000 ppm | |
CO₂ | 0 - 10 ... 20 % | |||
Other Measurement scales | NH₃ | 0 - 15 / 0 - 5000 ppm | ||
Dust | 0 - 10 / 0 - 1000 mg/Nm³ of water | |||
Pressure | 800 - 1200 mbar | |||
O₂ | 0 - 10 ... 25 % | |||
HCl | 0 - 10 / 0 - 5000 ppm | |||
Speed | 3 - 30 m/s (other datasheet optional) | |||
Temperature | 0 - 300 °C | |||
Repeatability | ≤ ± 0.5% PE (NDIR and ZrO2) | |||
≤ ± 1% PE (laser) | ||||
Linearity | ≤ ± 1 % PE | |||
Zero drift | ≤ ± 2% PE per week (NDIR) | ≤ ± 1% PE per week (NDIR) | ||
≤ ± 2% PE per month (ZrO₂) | ≤ ± 2% PE per month (ZrO₂) | |||
≤ ± 2% PE every 6 months (laser) | ≤ ± 2% PE every 6 months (laser) | |||
Scale drift | ≤ ± 2% PE per week (NDIR) | |||
≤ ± 2% PE per month (ZrO₂) | ||||
≤ ± 2% PE every 6 months (laser) |
Technologies
Continuous Emission Monitoring System CEMS allows the continuous monitoring of gas emissions in flue gas.
Gain precise insight into up to 14 gas components simultaneously with Fuji Electric gas analysis systems.
DAHS data acquisition and processing system - Your software for collecting, analyzing, reporting and generating statistical control charts for continuous emission monitoring (CEMS).